EpochXP Software Update History
EpochXP ver. 3.0.303 Neurological Monitoring Software update.
Latest update patch (Build 3.0.303)
New Features and fixes:
- Fixed saving screen snapshots. (Problem introduced in ver. 3.0302)
Patch History
3.0.302
- Added support for laptop computer built-in lameras.
- Fixed Controller communication protocol error (introduced in version 3.0.301).
- Fixed potential computer hang when Print button is clicked when memory resources are low.
3.0.301
- Added ability to superimpose trends.
- EMG threshold tone is now based on free run EMG and not Sig Trig trace.
- Fix Sig Trig traces of different sweep lengths in different Sets based on Common Set EMG
__________________________________________________________________________
EpochXP Ver. 2.3 Update History (Release Date 4/30/09)
EpochXP ver. 2.3.248 Neurological Monitoring Software update.
EpochXP ver. 2.3.242 Remote Reader Software update.
Latest update patch (Build 2.3.248)
New Features and fixes:
- Fixed Controller communication protocol error (introduced in version 2.3.247)
Patch History
2.3.247
- Set name is now appended to the "Snapshot save" comment.
- Changed font size for markers in the Stack report from 7 to 9 point.
- Disabled auto impedance checking during a single stimulus sequence.
- Display current sweep and building average on the remote review site.
- Continue to display EEG/EMG when the end of running test is reached.
- Fix crash when showing tooltip for interpeak markers when only the interpeak name is entered.
- Fix Sig Trig traces of different sweep lengths in the same Set based on Common Set EMG.
2.3.244
- Fixed stimulus Signal Trigger trace error when not all Sig Trig traces in the Common Set have stimulus names assigned.
- Fixed error message and restart procedure for SS311 and SS211 stimulators
2.3.242
- Fixed printing EMG Trace Area in Review.
- Fixed printing trend values and possible incorrect EEG Trace names in Export ASCII .
- Disabled audio masking noise when audio stimulus is switched Off.
2.3.241
- Message is displayed when the free disk space is below 10%.
- Message is displayed when the free disk space is below 1%.
- Electrode impedance is not checked when ESU is detected using the mute probe.
- Set Delay and Set Sequence limit schanged from 9999 to 99999 seconds.
- Updated Help file.
- Updated User Manual - ver 2.3 rev.2.
- Corrected measured current value in trace area when Single Sequence stimulation is used.
- Marker window is now displayed after latest Windows updates.
- Saving video now working correctly when video-to-USB convertor is plugged in without camera.
- Patient info is now saved when the patient name is hidden and the patient list is sorted when copying to a CD.
- Removed possible occurrence of a delay starting SigTrig traces in a new Set.
- The last trend value is now updated when the marker on the last average or baseline is changed.
- Fixed stimulation stop operation when "Stop Stimulation On Set Change" is ON and "Pause Stim On Stop" is OFF.
- Fixed displaying of some stack pages when the stack uses the maximum screen area.
- EP traces are now placed in the stack window when the number of sweeps is 1 and traces are collected faster than twice per second.
2.3.237
- Fixed installation problem.
2.3.236
- Added ability to change High Level Electrical Output Site during the test and added dialog box with choice to reset intensity to zero or retain last intensity.
- Added autoscrolling of long comment entry fields.
- Improved electrode impedance testingafter ESU.
- Changed minimum value of Suppress Stimulus Artifact time to1 ms.
- Fixed missing data line in Numeric Trend Report when page breaks.
- Fixed missing patient names and IDs in Patient Data List when list is sorted.
- Fixed problem when entering interpeak name before selecting interpeak markers
2.3.234
- Enabled Single Sequence mode for MEP stimulation.
- The size of the selected patient data is now displayed in MB, instead of KB.
- Included the video data file size to the total size of selected patient data.
2.3.233
- Changed default stimulation rate for single stimulus in a Set to 5 Hz.
- Fixed switching EEG traces OFF/ON when using TCeMEP stimulation.
- Fixed Error using two SS311 stimulators in different Sets.
- Improved synchronization of data and video.
2.3.231
- Fixed updating of Signal Triggered Stack when the last trace in the EMG trace area is switched off.
- Fixed playback of Signal Triggered data in review.
- Fixed ASCII export of trend data.
- Fixed problem associated with ASCII export dialog box.
|