Eclipse Software Update History
Eclipse Ver. 3.2.325 System full version Released 3/23/2010
Eclipse Ver. 3.2.325 Reader full version Released 3/23/2010
Key New Features, Changes and Fixes
Neurophysiologist Supported System:
- Added Nerve Proximity stimulation mode.
- Windows 7 compatibility.
- Live IOM remote monitoring (3 second delay).
- Added NIOM Explorer VPN browser providing status and control of all monitoring systems on the network.
- Added chat window when remote monitoring is implemented via Shared Folder.
- Added 5 s/div EMG timebase.
- Added Low Electrical Stim beep tone to indicate when measured current is first detected confirming tissue contact.
- Added ability to swap Active and Reference electrode columns in channel table.
- Improved sound quality of EMG and SigTrig tones.
- Unique EMG tones can be used to discern left and right EMG trace activity.
- Last Average acquisition time is now displayed during Runtime.
- Fixed mouse wheel interface when using the stimulus intensity entry field.
- Improved manual Set switching speed.
- Fast rebuild of EP traces in review.
- Added support for Invos 5100C Cerebral Oximeter.
- Modified factory support via the Internet.
Patch History
3.0.303
- Fixed Saving of screen snapshots. (Problem introduced in ver. 3.0.302)
3.0.302
- Added support for laptop computer built-in cameras.
- Fixed potential hang when the Print button is clicked and memory resources are low.
3.0.301
- Added auto-reset feature to controller during power dip or communication error.
- Added ability to superimpose trends.
- EMG threshold tone is now based on free run EMG and not Sig Trig trace.
- Fix Sig Trig traces of different sweep lengths in different Sets based on Common Set EMG.
- Fix USB controller/computer communication protocol error.
___________________________________________________________________________
Eclipse Ver. 2.3 Update History
Eclipse ver. 2.3.247 Neurological Monitoring Software update.
Eclipse ver. 2.3.242 Remote Reader Software update.
Latest update patch (Build 2.3.247)
New Features and fixes:
- Added auto-reset feature to controller during power dip or communication error.
- Fix Sig Trig traces of different sweep lengths in different Sets based on Common Set EMG.
- Fix USB controller/computer communication protocol error.
Patch History
2.3.246
- Set name is now appended to the "Snapshot save" comment.
- Changed font size for markers in the Stack report from 7 to 9 point.
- Improved USB controller/computer driver.
- Disabled auto impedance checking during a single stimulus sequence.
- Display current sweep and building average on the remote review site.
- Continue to display EEG/EMG when the end of running test is reached.
- Fix crash when showing tooltip for interpeak markers when only the interpeak name is entered.
2.3.244
- Fixed stimulus Signal Trigger trace error when not all Sig Trig traces in the Common Set have stimulus names assigned.
- Fixed potential increased noise problem when not all channels are used.
- Fixed measured current reading error after stimulus error.
2.3.242
- Fixed printing EMG Trace Area in Review.
- Fixed printing trend values and possible incorrect EEG Trace names in Export ASCII .
- Fixed possible Stimulator Error when interleaving audio and electrical stimuli.
- Disabled audio masking noise when audio stimulus is switched Off.
2.3.241
- Message is displayed when the free disk space is below 10%.
- Message is displayed when the free disk space is below 1%.
- Electrode impedance is not checked when ESU is detected using the mute probe.
- Set Delay and Set Sequence limit schanged from 9999 to 99999 seconds.
- Updated Help file.
- Updated User Manual - ver 2.3 rev.2.
- Updated Service Manual - rev. 2.1.
- Corrected measured current value in trace area when Single Sequence stimulation is used.
- Marker window is now displayed after latest Windows updates.
- Saving video now working correctly when video-to-USB convertor is plugged in without camera.
- Patient info is now saved when the patient name is hidden and the patient list is sorted when copying to a CD.
- Removed possible occurrence of a delay starting SigTrig traces in a new Set.
- The last trend value is now updated when the marker on the last average or baseline is changed.
- Accepted/Rejected numbers now are not reset when a trace is switched OFF or ON.
- Fixed possible "Controller disconnected" error message when Start or NextAvg is pressed after a Set change.
- Fixed stimulation stop operation when "Stop Stimulation On Set Change" is ON and "Pause Stim On Stop" is OFF.
- Fixed displaying of some stack pages when the stack uses the maximum screen area.
- EP traces are now placed in the stack window when the number of sweeps is 1 and traces are collected faster than twice per second.
2.3.237
- Fixed installation problem.
2.3.236
- Added ability to reconnect the Eclipse Controller during the test.
- Added ability to change High Level Electrical Output Site during the test and added dialog box with choice to reset intensity to zero or retain last intensity.
- Added autoscrolling of long comment entry fields.
- Added stop stimulation when muting is detected from ESU Mute Probe or Mute Trace.
- Improved electrode impedance testingafter ESU.
- Improved muting response time when Mute Trace is used.
- Changed minimum value of Suppress Stimulus Artifact time to1 ms.
- Changed minimum Train Count in Double Train mode to1.
- Fixed possible problem after Electrical Stimulator error message is displayed.
- Fixed missing data line in Numeric Trend Report when page breaks.
- Fixed missing patient names and IDs in Patient Data List when list is sorted.
- Fixed problem when entering interpeak name before selecting interpeak markers
2.3.234
- Enabled Single Sequence mode for MEP stimulation.
- The size of the selected patient data is now displayed in MB, instead of KB.
- Included the video data file size to the total size of selected patient data..
- Changed electrical stimulator self test voltage mode offset measuring limits.
- Fixed Standard Test use with double train stimulation created before version 2.3.
2.3.233
- Added support for electrical stimulator firmware version 5.0.
- Changed default stimulation rate for a single stimulus in a Set to 5 Hz .
- Fixed switching EEG traces OFF/ON when using TCeMEP stimulation.
- Fixed program crash opening the Speaker page when a trace name is too long
- Fixed electrical stimulator self test error using a Random Stimulation rate
- Improved synchronization of data and video
2.3.232
- Fixed potential electrical stimulator seft-test error.
2.3.231
- Fixed updating of Signal Triggered Stack when the last trace in the EMG trace area is switched off.
- Fixed playback of Signal Triggered data in review.
- Fixed ASCII export of trend data.
- Fixed problem associated with ASCII export dialog box.
|