Eclipse Ver. 3.2.325 System full version Released 3/23/2010

Eclipse Ver. 3.2.325 Reader full version Released 3/23/2010

Key New Features, Changes and Fixes

Neurophysiologist Supported System:

  • Added Nerve Proximity stimulation mode.
  • Windows 7 compatibility.
  • Live IOM remote monitoring (3 second delay).
  • Added NIOM Explorer VPN browser providing status and control of all monitoring systems on the network.
  • Added chat window when remote monitoring is implemented via Shared Folder.
  • Added 5 s/div EMG timebase.
  • Added Low Electrical Stim beep tone to indicate when measured current is first detected confirming tissue contact.
  • Added ability to swap Active and Reference electrode columns in channel table.
  • Improved sound quality of EMG and SigTrig tones.
  • Unique EMG tones can be used to discern left and right EMG trace activity.
  • Last Average acquisition time is now displayed during Runtime.
  • Fixed mouse wheel interface when using the stimulus intensity entry field.
  • Improved manual Set switching speed.
  • Fast rebuild of EP traces in review.
  • Added support for Invos 5100C Cerebral Oximeter.
  • Modified factory support via the Internet.

Patch History

3.0.303

  • Fixed Saving of screen snapshots. (Problem introduced in ver. 3.0.302)

3.0.302

  • Added support for laptop computer built-in cameras.
  • Fixed potential hang when the Print button is clicked and memory resources are low.

3.0.301

  • Added auto-reset feature to controller during power dip or communication error.
  • Added ability to superimpose trends.
  • EMG threshold tone is now based on free run EMG and not Sig Trig trace.
  • Fix Sig Trig traces of different sweep lengths in different Sets based on Common Set EMG.
  • Fix USB controller/computer communication protocol error.

___________________________________________________________________________

Eclipse Ver. 2.3 Update History

Eclipse ver. 2.3.247 Neurological Monitoring Software update.
Eclipse ver. 2.3.242 Remote Reader Software update.

Latest update patch (Build 2.3.247)

New Features and fixes:

  • Added auto-reset feature to controller during power dip or communication error.
  • Fix Sig Trig traces of different sweep lengths in different Sets based on Common Set EMG.
  • Fix USB controller/computer communication protocol error.

Patch History

2.3.246

  • Set name is now appended to the "Snapshot save" comment.
  • Changed font size for markers in the Stack report from 7 to 9 point.
  • Improved USB controller/computer driver.
  • Disabled auto impedance checking during a single stimulus sequence.
  • Display current sweep and building average on the remote review site.
  • Continue to display EEG/EMG when the end of running test is reached.
  • Fix crash when showing tooltip for interpeak markers when only the interpeak name is entered.

2.3.244

  • Fixed stimulus Signal Trigger trace error when not all Sig Trig traces in the Common Set have stimulus names assigned.
  • Fixed potential increased noise problem when not all channels are used.
  • Fixed measured current reading error after stimulus error.

2.3.242

  • Fixed printing EMG Trace Area in Review.
  • Fixed printing trend values and possible incorrect EEG Trace names in Export ASCII .
  • Fixed possible Stimulator Error when interleaving audio and electrical stimuli.
  • Disabled audio masking noise when audio stimulus is switched Off.

2.3.241

  • Message is displayed when the free disk space is below 10%.
  • Message is displayed when the free disk space is below 1%.
  • Electrode impedance is not checked when ESU is detected using the mute probe.
  • Set Delay and Set Sequence limit schanged from 9999 to 99999 seconds.
  • Updated Help file.
  • Updated User Manual - ver 2.3 rev.2.
  • Updated Service Manual - rev. 2.1.
  • Corrected measured current value in trace area when Single Sequence stimulation is used.
  • Marker window is now displayed after latest Windows updates.
  • Saving video now working correctly when video-to-USB convertor is plugged in without camera.
  • Patient info is now saved when the patient name is hidden and the patient list is sorted when copying to a CD.
  • Removed possible occurrence of a delay starting SigTrig traces in a new Set.
  • The last trend value is now updated when the marker on the last average or baseline is changed.
  • Accepted/Rejected numbers now are not reset when a trace is switched OFF or ON.
  • Fixed possible "Controller disconnected" error message when Start or NextAvg is pressed after a Set change.
  • Fixed stimulation stop operation when "Stop Stimulation On Set Change" is ON and "Pause Stim On Stop" is OFF.
  • Fixed displaying of some stack pages when the stack uses the maximum screen area.
  • EP traces are now placed in the stack window when the number of sweeps is 1 and traces are collected faster than twice per second.

2.3.237

  • Fixed installation problem.

2.3.236

  • Added ability to reconnect the Eclipse Controller during the test.
  • Added ability to change High Level Electrical Output Site during the test and added dialog box with choice to reset intensity to zero or retain last intensity.
  • Added autoscrolling of long comment entry fields.
  • Added stop stimulation when muting is detected from ESU Mute Probe or Mute Trace.
  • Improved electrode impedance testingafter ESU.
  • Improved muting response time when Mute Trace is used.
  • Changed minimum value of Suppress Stimulus Artifact time to1 ms.
  • Changed minimum Train Count in Double Train mode to1.
  • Fixed possible problem after Electrical Stimulator error message is displayed.
  • Fixed missing data line in Numeric Trend Report when page breaks.
  • Fixed missing patient names and IDs in Patient Data List when list is sorted.
  • Fixed problem when entering interpeak name before selecting interpeak markers

2.3.234

  • Enabled Single Sequence mode for MEP stimulation.
  • The size of the selected patient data is now displayed in MB, instead of KB.
  • Included the video data file size to the total size of selected patient data..
  • Changed electrical stimulator self test voltage mode offset measuring limits.
  • Fixed Standard Test use with double train stimulation created before version 2.3.

2.3.233

  • Added support for electrical stimulator firmware version 5.0.
  • Changed default stimulation rate for a single stimulus in a Set to 5 Hz .
  • Fixed switching EEG traces OFF/ON when using TCeMEP stimulation.
  • Fixed program crash opening the Speaker page when a trace name is too long
  • Fixed electrical stimulator self test error using a Random Stimulation rate
  • Improved synchronization of data and video

2.3.232

  • Fixed potential electrical stimulator seft-test error.

2.3.231

  • Fixed updating of Signal Triggered Stack when the last trace in the EMG trace area is switched off.
  • Fixed playback of Signal Triggered data in review.
  • Fixed ASCII export of trend data.
  • Fixed problem associated with ASCII export dialog box.

   
 
©2009-Axon Systems, Inc. All rights reserved.