Eclipse Software Update Patches
Software update patches resolve bugs or add new features associated with Eclipse Neurological Workstation and Remote Reader software versions. Read Installation Notes
Eclipse Ver. 2.3 Update Patches (Release Date 3/21/08)
Download Eclipse ver. 2.3.244 Neurological Monitoring Software update.
Download Eclipse ver. 2.3.242 Remote Reader Software update.
Latest update patch (Build 2.3.244)
New Features and fixes:
- Fixed stimulus Signal Trigger trace error when not all Sig Trig traces in the Common Set have stimulus names assigned.
- Fixed potential increased noise problem when not all channels are used.
- Fixed measured current reading error after stimulus error.
Patch History
2.3.242
- Fixed printing EMG Trace Area in Review.
- Fixed printing trend values and possible incorrect EEG Trace names in Export ASCII .
- Fixed possible Stimulator Error when interleaving audio and electrical stimuli.
- Disabled audio masking noise when audio stimulus is switched Off.
2.3.241
- Message is displayed when the free disk space is below 10%.
- Message is displayed when the free disk space is below 1%.
- Electrode impedance is not checked when ESU is detected using the mute probe.
- Set Delay and Set Sequence limit schanged from 9999 to 99999 seconds.
- Updated Help file.
- Updated User Manual - ver 2.3 rev.2.
- Updated Service Manual - rev. 2.1.
- Corrected measured current value in trace area when Single Sequence stimulation is used.
- Marker window is now displayed after latest Windows updates.
- Saving video now working correctly when video-to-USB convertor is plugged in without camera.
- Patient info is now saved when the patient name is hidden and the patient list is sorted when copying to a CD.
- Removed possible occurrence of a delay starting SigTrig traces in a new Set.
- The last trend value is now updated when the marker on the last average or baseline is changed.
- Accepted/Rejected numbers now are not reset when a trace is switched OFF or ON.
- Fixed possible "Controller disconnected" error message when Start or NextAvg is pressed after a Set change.
- Fixed stimulation stop operation when "Stop Stimulation On Set Change" is ON and "Pause Stim On Stop" is OFF.
- Fixed displaying of some stack pages when the stack uses the maximum screen area.
- EP traces are now placed in the stack window when the number of sweeps is 1 and traces are collected faster than twice per second.
2.3.237
- Fixed installation problem.
2.3.236
- Added ability to reconnect the Eclipse Controller during the test.
- Added ability to change High Level Electrical Output Site during the test and added dialog box with choice to reset intensity to zero or retain last intensity.
- Added autoscrolling of long comment entry fields.
- Added stop stimulation when muting is detected from ESU Mute Probe or Mute Trace.
- Improved electrode impedance testingafter ESU.
- Improved muting response time when Mute Trace is used.
- Changed minimum value of Suppress Stimulus Artifact time to1 ms.
- Changed minimum Train Count in Double Train mode to1.
- Fixed possible problem after Electrical Stimulator error message is displayed.
- Fixed missing data line in Numeric Trend Report when page breaks.
- Fixed missing patient names and IDs in Patient Data List when list is sorted.
- Fixed problem when entering interpeak name before selecting interpeak markers
2.3.234
- Enabled Single Sequence mode for MEP stimulation.
- The size of the selected patient data is now displayed in MB, instead of KB.
- Included the video data file size to the total size of selected patient data..
- Changed electrical stimulator self test voltage mode offset measuring limits.
- Fixed Standard Test use with double train stimulation created before version 2.3.
2.3.233
- Added support for electrical stimulator firmware version 5.0.
- Changed default stimulation rate for a single stimulus in a Set to 5 Hz .
- Fixed switching EEG traces OFF/ON when using TCeMEP stimulation.
- Fixed program crash opening the Speaker page when a trace name is too long
- Fixed electrical stimulator self test error using a Random Stimulation rate
- Improved synchronization of data and video
2.3.232
- Fixed potential electrical stimulator seft-test error.
2.3.231
- Fixed updating of Signal Triggered Stack when the last trace in the EMG trace area is switched off.
- Fixed playback of Signal Triggered data in review.
- Fixed ASCII export of trend data.
- Fixed problem associated with ASCII export dialog box.
Eclipse Ver. 2.0 Update Patches (Release Date 11/15/06)
Download Eclipse ver. 2.00.210 Neurological Monitoring Software update.
Download Eclipse ver. 2.00.209 Remote Reader Software update.
Latest update patch (Build 2.00.210)
New Features and fixes:
- Added support for electrical stimulator versions 4.2 and 4.3.
- Fixed patient name not displayed in Patient List Report.
- Fixed Impedance checking in Low Electrical stimulator voltage mode.
Patch History
2.00.209
- Upgraded electrode table to account for channel one not used in test.
- Fixed initialization of encryption object to allow loading on all computers.
2.00.208
- Upgraded to latest release of remote monitoring/support software application.
- Changed Pedicle Integity Test default Start Time from 0 to 6 ms.
- Improved Pedicle Integrity Test probe fault detection.
- Reduced crosstalk between channels when input voltage range is exceeded.
- Changed minimum time between sweeps from 5 to 8 ms.
2.00.207
- Fixes electrical stimulus intensity problem : When interleaving at the fastest stimulation rate, the current delivered to the patient could be less than the set current.
2.00.206
- Added ability to set and delete markers in the Stack Window.
- Click on a stack trace to quickly position the cursor at that trace.
- Cursor is automatically positioned at a selected (highlighted) stack trace.
- Artifact Level and Artifact Start entry fields are disabled when the Artifact check box is cleared.
- Added Active and Reference labels to + and - channel input labels.
- Marker latency is correctly displayed in Trace Window when Autotracking is OFF.
- Markers moved in the Trace Window are correctly updated in the Stack window.
- Corrected Suppress Stim artifact error for Digital Preamplifier B.
- Set name and stimulation rate now correctly saved in a Set with Signal Triggered traces only.
- Improved audio muting of Stim Artifact .
- Latency of Stack markers now correctly positioned.
- Comment text entered during a Set change is preserved.
- Electrode impedances are now measured for Digital Preamplifier B even if not used in the current Set.
2.00.205
- Changed font point size and layout of Impedance Report.
- Selected video source saved when exiting test.
- Last Average trace correctly shown or hidden based on LastAvg checkmark setting.
- Corrected highlighting of selected traces in split Trace Area.
- Max train rate for Fast Charge TCeMEP increased from 500Hz to 1000Hz
2.00.204
- ""Impedance Comments" checkmark in Impedance Window automatically inserts impedance values in Comments History.
- Last average checkmark now shows or hides Last Average in all averaging modes.
- Marker mode settings are saved with system.
- Time now displayed for all stack traces.
- Standard Test name length changed from 70 to 43 characters.
- Trace Area text positions changed to improve readability.
- EP Markers Building Average Checkbox now shows or hides Building Average when EP Marker Window is open.
- Clicking on Baseline, Building or Last Average trace now selects Marker mode
- Selecting a time point on Building Average sets Autotracking parameters.
- Printing now uses modified print preferences.
- Increased speed of snapshot saves.
- Decreased refesh time of run-time Stack. .
- Fixed corrupted screen layout with video and Set change.
- Fixed reviewing of screens from CD drive.
- Fixed video review when video is not saved from the beginning of the test.
Installation Notes:
- The latest patch includes all the changes incorporated in the previous patches.
- The update patches are for use ONLY with the complete Eclipse Neurological Monitoring or Remote Reader software beginning with version 2.00. The patches will NOT install the complete application.
- Download the desired update. The update can be installed automatically or saved to your hard drive for installation at a later time. If the patch is saved to a folder, click on the update icon to automatically install the patch. The patch file size is approximately 14MB.
- If you experience any problems please contact Axon Systems customer support.
|